Building on 50 years of Philips experience, the HeartStart XL+ is designed for Resuscitation and Rapid Response Teams. It contains meaningful innovations that can help you confidently and effectively respond to patients across the hospital.
Humidity | Up to 95
% relative humidity
|
Atmospheric pressure range | Operating and storage - 1014 mbar to 572 mbar(0 to 15000 ft; 0 to 4500 m)
|
Shock(a) | Operating Half-sine waveform/duration ≤11 ms/acceleration ≥ 15.3 G at 3 shocks per face
|
Shock(b) | Non-operating Trapezoidal waveform/acceleration 30G/ velocity change 7.42 m/s +/-10% 1 shock per face
|
Water/solids ingress resistance | Meets Ingress Protection level IP21
|
EMC | Complies with the requirements of standard EN 60601-1-2:2002
|
Safety | Meets UL 60601-1 (1st edition)/EN 60601-2-4:2003/EN 60601-1:1990
|
Mode of operation | Continuous
|
AC Line powered | 100-240 VAC/50 or 60 Hz/1-0.46A/Class I Equipment
|
Battery powered | Minimum 14.4 V rechargeable lithium ion VAC/Hz
|
Size | Size 9˝ high x 11.6˝ wide x 10.9˝ deep (23 cm x 29.6 cm x 27.9 cm)
|
Type | Rechargeable, lithium ion; see battery label for capacity information
|
Dimensions | 1˝ high x 4.5˝ wide x 5.7˝ long (23.6 mm high x 116 mm wide x 146 mm long)
|
Capacity | With a new fully charged battery, at 20 °C (68 °F) one of the following: At least 3 hours of monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks. OR at least two hours of pacing (180ppm at 140mA with 40 msec pulse width) while monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks; OR at least 175 full energy charge/shocks
|
Weight | Approximately 1.5 lbs (0.68kg)
|
Charge time device off | With the temperature between 0-35° C (32-95° F) less than 3 hours to 100% capacity and less than 2 hours to 80% capacity
|
Life | Approximately 3
yr
|
Battery indicators | Battery gauge on battery, capacity indicator on display, power indicators on front of device; flashing RFU indicator, chirp and Low Battery messages on the display for low battery condition. When a low battery message first appears there is still enough energy for at least 10 minutes of monitoring and 6 maximum energy discharges
|
Battery storage | Storing the battery for extended periods at temperatures above 40° C (104° F) reduces battery capacity and degrades battery life
|
Weight | Weight 14.7 lbs (6.6 kg), includes one battery/one new roll of paper/one pads cable. Incremental weight of external standard paddles and paddle tray is less than 3 lbs (1.3 kg)
|
Standard operator position | Within one meter (3 feet) of the device
|
Type | Rechargeable lithium ion battery; AC power using a protectively grounded outlet
|
Size | Approximately 6.5" (16.5 cm) diagonal viewing area
|
Type | Approximately 6.5" (16.5 cm) diagonal viewing area
|
Resolution | 640 x 480 pixels (VGA) with 32 brightness levels per color
|
Sweep speed | 20 mm/s nominal (stationary trace; sweeping erase bar) for ECG and SpO2
mm/s
|
Wave viewing time | 5.2
sec
|
Operating temperature | 0°C to 45°C (32°F to 113°F)
|
Storage temperature | 20°C to 70°C (-4°F to 158°F)
|
Waveform | Biphasic Truncated Exponential. Waveform parameters adjusted as a function of patient impedance
|
|||
Shock delivery | Via multifunction electrode pads or paddles
|
|||
Charge time | 3 seconds to the recommended adult energy level (150 Joules) with a new fully-charged battery installed
|
Less than 5 seconds to the selected energy level (up to 200 Joules) with a new fully charged battery installed
|
The device powers on in manual defibrillation mode ready to deliver shock in less than 8 seconds plus applicable charge time assuming an immediate selection of an energy and initiation of a charge even at 90V AC and after 15 maximum energy discharges
|
The device powers on in AED mode ready to deliver shock in less than 17 seconds plus applicable charge time
|
Shock series | Configurable energy escalation in a series
|
|||
Leads sensing | Leads off sensing and PCI sensing for pads/paddles. Apply 500nA rms (571Hz); 200uA rms (32KHz)
|
|||
Minimum patient impedance range | 25 ohm (external defibrillation);15 ohm (internal defibrillation)
ohm
|
|||
Maximum patient impedance range | 250 ohm. Actual functional range may exceed these values
ohm
|
Continuous ECG strip | The Print button starts and stops the strip. The printer can be configured to be run real time or with a 10-second delay. The strip prints the primary ECG lead and a second wave with event annotations and measurements
|
Auto printing | The printer can be configured to automatically print on Mark Events, Charge, Shock, and Alarm
|
Reports | The following can be printed: Event Summary (Long or Short)/Vital Signs Trends/Operational Check/Configuration/Status Log/Device Information
|
Speed | 25 mm/s with an accuracy of +/-5% mm/s
|
Amplitude accuracy | 5% for offset voltages of +/- 300 mV at 5Hz
|
Paper size | 50 mm wide x 30 m long
|